Most MedTech founders ask the pathway question too late. The prototype works, the pitch deck says "510(k) in 18 months," and only then does someone search for a predicate and find nothing that fits. A sound FDA 510(k) strategy starts much earlier, because the choice between a 510(k) and a De Novo request shapes your testing plan, your budget and the story you tell investors.
The stakes are real. For fiscal year 2027, the FDA's standard user fee is $28,653 for a 510(k) and $191,020 for a De Novo request, with reduced fees of $7,163 and $47,755 for qualifying small businesses (Federal Register, July 30, 2026). The fee is the smallest part of the difference. The evidence each pathway asks for is the larger one.
Here is how we help device teams think through the decision, and what has changed in 2026.
510(k) vs De Novo: the short version
A 510(k) asks FDA to agree that your device is substantially equivalent to a legally marketed predicate: same intended use, and either the same technological characteristics or differences that do not raise new questions of safety and effectiveness. You borrow a path someone else already cleared.
De Novo is for novel low-to-moderate-risk devices with no suitable predicate. FDA classifies the device into Class I or II and sets the general and special controls that will apply to it. You can file a De Novo directly if you conclude no predicate exists, or after a 510(k) comes back "not substantially equivalent." FDA's stated goal is a decision within 150 review days, De Novo requests must be filed through eSTAR, and a granted De Novo can become the predicate for later 510(k)s of that device type (FDA).
That last point matters for founders. A De Novo costs more up front, but it can create the category your competitors will have to cite.
An FDA 510(k) strategy decision guide: six questions
We walk teams through these questions before anyone commits to a regulatory timeline in a pitch deck.
1. Can you write your intended use in one sentence?
Pathway decisions start with the indications for use, not the hardware. Write the patient population, the clinical setting, the user and what the device does. If the sentence keeps growing, the pathway will be hard to pin down. Narrowing the first indication is often the fastest route to market, with expansions to follow.
2. Is there a predicate with the same intended use?
Search the FDA 510(k) and De Novo databases by product code, then read the decision summaries, not only the device names. Shortlist two or three candidates. A predicate that matches your intended use but differs in technology can still work. One that matches your technology but not your intended use usually will not.
3. Do your technological differences raise new questions?
This is where most 510(k) strategies succeed or fail. A new energy source, a new patient-contacting material, a new software function or a new anatomical location can each push a device out of substantial equivalence. List every difference from your lead predicate and ask what test would answer the question each one raises. If no standard bench test can answer it, plan for the evidence a De Novo would need.
4. What evidence will each pathway require?
Sketch both evidence plans side by side: bench and performance testing, biocompatibility, sterilization and shelf life, software documentation, electrical safety, human factors, animal or cadaver work, and any clinical data. Then put rough costs and months against each. The pathway with the lower user fee is not always the cheaper program once testing is counted.
5. How does the pathway fit your raise?
Investors fund milestones. A 510(k) with a clean predicate is a shorter, more predictable story. A De Novo is a bigger bet that can create a defensible category. Either can be the right answer, but the capital plan has to match the evidence plan, with room for an additional-information request in the middle of review.
6. What do you need FDA to confirm before you test?
FDA recommends a Pre-Submission before a De Novo request, and a pre-sub is worth considering for a 510(k) with a stretched predicate. Bring specific questions: Is this predicate appropriate? Is this test method acceptable? Is this sample size reasonable? Vague questions get vague answers.
What changed in 2026
Faster review goals are on the way. The draft MDUFA VI agreement, shared July 8, 2026, would cut the 510(k) review goal from 128 to 112 days over fiscal years 2028 to 2032, and would add Focused Follow-Up Pre-Submissions with a written FDA response within 45 days by the end of 2027 (MedTech Dive). Goals are not guarantees, so plan around today's process.
De Novo pain points are on the table. In the MDUFA VI talks, industry pointed to a lower success rate and more withdrawals for De Novo requests than for other submissions, and to limited options when a decision goes the wrong way (RAPS). That is a reason to invest in the pre-sub and the evidence plan, not a reason to avoid the pathway.
Quality data in submissions is an FDA priority. CDRH placed final guidance on quality management system information in premarket submissions on its top-priority list for fiscal year 2027 (Emergo by UL). With QMSR in force, your design controls are part of the submission story.
A 510(k) strategy does not end at clearance. Recent FDA warning letters cited material and formulation changes made to cleared devices without a new 510(k) (FDA Law Blog, Nov 11, 2026). Supplier and material decisions made during scale-up need a regulatory check.
Where we see teams get stuck
Picking the predicate after the design is frozen. By then, the differences are locked in and the test plan has to stretch to cover them.
Predicate creep. Chaining features from several predicates to build a device none of them resemble is a common route to a "not substantially equivalent" letter.
Treating the fee as the cost. The testing program, the months of review and the engineering time to answer questions usually dwarf the user fee on either path.
Skipping the pre-sub to save time. A few months for FDA feedback is often cheaper than repeating a test FDA does not accept.
Forgetting manufacturing. Materials, sterilization method and packaging chosen for the submission become the baseline you will have to defend every time production changes.
How Birch Design helps
Birch Design's FDA strategy work covers pathway and predicate strategy, Pre-Submission (Q-Sub) packages and an evidence plan aligned to a 510(k), De Novo or PMA. We have led devices through FDA pathways including preclinical studies, animal trials, first-in-human programs and 510(k)s, across areas like urology, orthopedics, diagnostics and catheter-based devices, so the regulatory plan is tied to the engineering and manufacturing work that has to support it.
If you are still weighing the pathway, many founders start with our Product Development & Commercialization Readiness Assessment: a prioritized gap list, a milestone roadmap with budget ranges and an investor-ready summary, with the regulatory pathway built in.
Start the conversation
Not sure whether your device is a 510(k) or a De Novo? Bring your intended use and your candidate predicates, and we will help you pressure-test the plan before you spend on testing. Contact Birch Design or book a discovery call.
Sources: FDA De Novo Classification Request page; Federal Register FY 2027 medical device user fee notice (July 30, 2026); MedTech Dive and RAPS coverage of MDUFA VI; Emergo by UL on CDRH FY 2027 guidance priorities; FDA Law Blog (Nov 11, 2026). This post is general information, not regulatory advice for a specific device.